TT7000 RF Power/Frequency Meter

$1,795.00

Available on back-order

Description

DS Instruments TT7000 RF Power/Frequency Meter

DS Instruments TT7000 (R6) – USB WIDEBAND RF METER / GENERATOR / COUNTER DS Instruments TT7000 (R6) – USB WIDEBAND RF METER / GENERATOR / COUNTER

TT7000, A Multi-Purpose Instrument

The DS Instruments TT7000 RF Power/Frequency Meter combines a Microwave Power Meter, Signal Source, Frequency counter, and RF Divider in one compact instrument. The frequency counter and power meter cover up to 7000MHz. The signal generator covers from 300 to 9600MHz.

Counting and Power reading can be done simultaneously over the entire band. The power meter dynamic range is from -50dBm to +5dBm. Accuracy is typically +/- 1.0dB from 0dBm to -30dBm when the frequency count is valid. Frequency can be manually set for low power signals to calibrate the power reading value.

The SG function is fully synthesized with step size less than 10Hz using precision internal or external 10MHz reference. Adjustable output power from +10dBm to -20dBm is now standard for the TT7000 with 0.25dB steps! Signal generator and power meter can function simultaneously and will auto-calibrate to the set frequency.

The RF input can also be processed by a RF Divider with 1/2/4/8 options. The output of the RF Divider is available on the RF OUT port with adjustable power level.

Front Panel and USB Interfaces

With the DS Instruments TT7000 RF Power/Frequency Meter connected to a host PC via its USB port, which is configured as a virtual COM port, industry standard SCPI text commands can control the TT7000 and retrieve instrument measurement data. The OLED display and interface buttons allow the user to not only verify incoming SCPI commands from the USB interface, but also control the unit in standalone mode with NO HOST PC REQUIRED.

Never exceed the maximum power of +15dBm on the meter INPUT port!

DS Instruments Battery Indication for signal level meters and tracking generators

  • Power measurements can be inaccurate if the automatic counter cannot determine an
    approximate frequency. Use the PC GUI to manually enter the frequency range of your signal if
    auto-range is not consistent.
  • If the power level is too low or there is no signal input, the frequency counter display will wonder.
    This is normal functionality.

DS Instruments Battery Indication for signal level meters and tracking generators

The TT7000 requires up to 0.75 Amperes in certain modes. Some legacy USB ports
are limited to 0.5 Amperes. A stable 5.0 volt input voltage is critical for reliable
operation within the unit s specifications.

Many common ways to satisfy this requirement exist:

  • A high quality USB cable connected directly to a POWERED USB hub.
  • A high quality dual-port USB cable connected to two USB ports.
  • A high quality USB cable plugged into a blue USB 3 port.
  • A lithium-ion USB battery pack
  • A USB smart phone charger (5 Volt wall adapter)

*A High Quality USB cable uses a heavier gauge wire resulting in less voltage loss from end to end.
Our suggested cable is maximum 1 meter length with 22 AWG conductors.

SPECIFICATIONS

TT7000 SIGNAL GENERATOR SPECIFICATIONS

Parameter Min Max Typ Units
Frequency Range (Adjustable power – Band 1) 300 4800 MHz
RF Output Power (Calibrated Mode – Band 1) -21 +10 dBm
RF Output Power (Uncalibrated Mode – Band 1) -35 +12 dBm
Frequency Range (Fixed – Band 2) 4800 9600 MHz
RF Output Power (Fixed: 4800-9600MHz Band 2) +10 +17 +13 dBm
 Phase Noise @ 9600MHz, 10KHz Offset -76 dBc
 @ 4800MHz, 10KHz Offset -80 dBc
 @ 2100MHz, 10KHz Offset -87 dBc
 @ 900MHz, 10KHz Offset -95 dBc
 @ 433MHz, 10KHz Offset -102 dBc
 @ 315MHz, 10KHz Offset -104 dBc
 Output Port Return Loss 6 10 dB
Frequency Step Size (Band 2 is doubled) 10 Hz
 Step Attenuator Range (Calibrated Band) 31.5 dB
 Attenuator Step Size (Calibrated Band) 0 0.25 dB
Variable Attenuator Dyn. Range (Calibrated Band) 15 dB
Power Output Level Typical Accuracy +/-1.5 dBm
USB Voltage 4.75 5.35 5.0 Vdc
Device Current Requirement 0.75 A
 Temperature Range -30 +50 25 C
Reference Frequency MCX Input (10MHz) -5 +12 +5 dBm
 Internal 10MHz Reference Stability +/-280 PPB

TT7000 FREQUENCY COUNTER SPECIFICATIONS

Conditions: 25C, USB supplied power

Parameter Min Max Typ Units
Input Frequency Range 100 7000 MHz
 Input Power Sensitivity -25 +10 0 dBm
Input Impedance 50 Ohm
Input VSWR, 50 to 7000MHz 1.3:1 1.6:1 1.5:1 SWR
 Internal 10MHz Time Base Stability +/-280 PPB
Resolution @ 1sec Gate Time 1000 Hz
 Gate Times
( Front Panel Operation Only Supports 1sec Gate) 0.1 10 1 Sec

TT7000 RMS POWER METER SPECIFICATIONS

Parameter Min Max Typ Units
Input Frequency Range 50 7000 MHz
Dynamic Range (0.01 – 2.0GHz) -54 +9 dBm
 Dynamic Range (2.0 – 6.0GHz) -45 +5 dBm
Dynamic Range (6.0 – 7.5GHz) -35 +5 dBm
Max Power Input (Damage Level) +15 dBm
Input Impedance 50 Ohm
 Input VSWR, 50 to 7000MHz 1.3:1 1.8:1 SWR
 Resolution 0.01 dBm
Absolute Power Accuracy +/-2.00 +/-0.75 dBm
Relative Power Accuracy +/-0.75 +/-0.25 dBm
Temperature Stability +/-0.25 dBm
Power Meter Architecture RMS
 Measurement Speed 25 500 100 mS
Averaging Range 1 50 10

TT7000 RF DIVIDER SPECIFICATIONS

Conditions: 25C

Parameter Min Max Typ Units
Input Frequency Range @ RF IN Jack 300 7000 MHz
Output Frequency Range @ RF OUT Jack 37.5 3500 MHz
Divide Ratios Available 1/2/4/8
 Input Power Sensitivity -10 +10 0 dBm
 Input Impedance @ RF IN Jack 50 Ohm
 Input VSWR, 300 to 7000MHz 1.2:1 1.6:1 1.4:1 SWR
 Output Power @ RF Out Jack +7 +14 +10 dBm
Output Impedance @ RF OUT Jack 50 Ohm
Output VSWR 1.3:1 1.6:1 1.25:1 SWR

 

DS Instruments TT7000 GUI DS Instruments TT7000 GUI

DS Instruments TT7000 Phase Noise DS Instruments TT7000 Phase Noise

DS Instruments TT7000 RF Power Meter and Frequency Counter Data Sheet