Description
DS Instruments TT7000 RF Power/Frequency Meter
TT7000, A Multi-Purpose Instrument
The DS Instruments TT7000 RF Power/Frequency Meter combines a Microwave Power Meter, Signal Source, Frequency counter, and RF Divider in one compact instrument. The frequency counter and power meter cover up to 7000MHz. The signal generator covers from 300 to 9600MHz.
Counting and Power reading can be done simultaneously over the entire band. The power meter dynamic range is from -50dBm to +5dBm. Accuracy is typically +/- 1.0dB from 0dBm to -30dBm when the frequency count is valid. Frequency can be manually set for low power signals to calibrate the power reading value.
The SG function is fully synthesized with step size less than 10Hz using precision internal or external 10MHz reference. Adjustable output power from +10dBm to -20dBm is now standard for the TT7000 with 0.25dB steps! Signal generator and power meter can function simultaneously and will auto-calibrate to the set frequency.
The RF input can also be processed by a RF Divider with 1/2/4/8 options. The output of the RF Divider is available on the RF OUT port with adjustable power level.
Front Panel and USB Interfaces
With the DS Instruments TT7000 RF Power/Frequency Meter connected to a host PC via its USB port, which is configured as a virtual COM port, industry standard SCPI text commands can control the TT7000 and retrieve instrument measurement data. The OLED display and interface buttons allow the user to not only verify incoming SCPI commands from the USB interface, but also control the unit in standalone mode with NO HOST PC REQUIRED.
Never exceed the maximum power of +15dBm on the meter INPUT port!
- Power measurements can be inaccurate if the automatic counter cannot determine an
approximate frequency. Use the PC GUI to manually enter the frequency range of your signal if
auto-range is not consistent. - If the power level is too low or there is no signal input, the frequency counter display will wonder.
This is normal functionality.
The TT7000 requires up to 0.75 Amperes in certain modes. Some legacy USB ports
are limited to 0.5 Amperes. A stable 5.0 volt input voltage is critical for reliable
operation within the unit s specifications.
Many common ways to satisfy this requirement exist:
- A high quality USB cable connected directly to a POWERED USB hub.
- A high quality dual-port USB cable connected to two USB ports.
- A high quality USB cable plugged into a blue USB 3 port.
- A lithium-ion USB battery pack
- A USB smart phone charger (5 Volt wall adapter)
*A High Quality USB cable uses a heavier gauge wire resulting in less voltage loss from end to end.
Our suggested cable is maximum 1 meter length with 22 AWG conductors.
SPECIFICATIONS
TT7000 SIGNAL GENERATOR SPECIFICATIONS
Parameter | Min | Max | Typ | Units | |
Frequency Range (Adjustable power – Band 1) | 300 | 4800 | MHz | ||
RF Output Power (Calibrated Mode – Band 1) | -21 | +10 | dBm | ||
RF Output Power (Uncalibrated Mode – Band 1) | -35 | +12 | dBm | ||
Frequency Range (Fixed – Band 2) | 4800 | 9600 | MHz | ||
RF Output Power (Fixed: 4800-9600MHz Band 2) | +10 | +17 | +13 | dBm | |
Phase Noise | @ 9600MHz, 10KHz Offset | -76 | dBc | ||
@ 4800MHz, 10KHz Offset | -80 | dBc | |||
@ 2100MHz, 10KHz Offset | -87 | dBc | |||
@ 900MHz, 10KHz Offset | -95 | dBc | |||
@ 433MHz, 10KHz Offset | -102 | dBc | |||
@ 315MHz, 10KHz Offset | -104 | dBc | |||
Output Port Return Loss | 6 | 10 | dB | ||
Frequency Step Size (Band 2 is doubled) | 10 | Hz | |||
Step Attenuator Range (Calibrated Band) | 31.5 | dB | |||
Attenuator Step Size (Calibrated Band) | 0 | 0.25 | dB | ||
Variable Attenuator Dyn. Range (Calibrated Band) | 15 | dB | |||
Power Output Level Typical Accuracy | +/-1.5 | dBm | |||
USB Voltage | 4.75 | 5.35 | 5.0 | Vdc | |
Device Current Requirement | 0.75 | A | |||
Temperature Range | -30 | +50 | 25 | C | |
Reference Frequency MCX Input (10MHz) | -5 | +12 | +5 | dBm | |
Internal 10MHz Reference Stability | +/-280 | PPB |
TT7000 FREQUENCY COUNTER SPECIFICATIONS
Conditions: 25C, USB supplied power
Parameter | Min | Max | Typ | Units | |
Input Frequency Range | 100 | 7000 | MHz | ||
Input Power Sensitivity | -25 | +10 | 0 | dBm | |
Input Impedance | 50 | Ohm | |||
Input VSWR, 50 to 7000MHz | 1.3:1 | 1.6:1 | 1.5:1 | SWR | |
Internal 10MHz Time Base Stability | +/-280 | PPB | |||
Resolution @ 1sec Gate Time | 1000 | Hz | |||
Gate Times | |||||
( Front Panel Operation Only Supports 1sec Gate) | 0.1 | 10 | 1 | Sec |
TT7000 RMS POWER METER SPECIFICATIONS
Parameter | Min | Max | Typ | Units | |
Input Frequency Range | 50 | 7000 | MHz | ||
Dynamic Range (0.01 – 2.0GHz) | -54 | +9 | dBm | ||
Dynamic Range (2.0 – 6.0GHz) | -45 | +5 | dBm | ||
Dynamic Range (6.0 – 7.5GHz) | -35 | +5 | dBm | ||
Max Power Input (Damage Level) | +15 | dBm | |||
Input Impedance | 50 | Ohm | |||
Input VSWR, 50 to 7000MHz | 1.3:1 | 1.8:1 | SWR | ||
Resolution | 0.01 | dBm | |||
Absolute Power Accuracy | +/-2.00 | +/-0.75 | dBm | ||
Relative Power Accuracy | +/-0.75 | +/-0.25 | dBm | ||
Temperature Stability | +/-0.25 | dBm | |||
Power Meter Architecture | RMS | ||||
Measurement Speed | 25 | 500 | 100 | mS | |
Averaging Range | 1 | 50 | 10 |
TT7000 RF DIVIDER SPECIFICATIONS
Conditions: 25C
Parameter | Min | Max | Typ | Units | |
Input Frequency Range @ RF IN Jack | 300 | 7000 | MHz | ||
Output Frequency Range @ RF OUT Jack | 37.5 | 3500 | MHz | ||
Divide Ratios Available | 1/2/4/8 | ||||
Input Power Sensitivity | -10 | +10 | 0 | dBm | |
Input Impedance @ RF IN Jack | 50 | Ohm | |||
Input VSWR, 300 to 7000MHz | 1.2:1 | 1.6:1 | 1.4:1 | SWR | |
Output Power @ RF Out Jack | +7 | +14 | +10 | dBm | |
Output Impedance @ RF OUT Jack | 50 | Ohm | |||
Output VSWR | 1.3:1 | 1.6:1 | 1.25:1 | SWR |